Fluke 5700A Series II Service Manual page 100

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II Calibrator
Service Manual
Use Table 3-23 (5720A) or Table 3-24 (5700A) for the test records.
1. Place the 5790A on top of the Calibrator and connect the equipment as shown in
The point of measurement is at the end of the cable and adapter that
connects to the 5790A. Other cable lengths and adapters will yield different
results at high frequencies.
2. On the 5790A push UTIL MENUS button and then the MEAS CONTROL softkey.
3. Set the 5790A to 2 mV at 1 kHz. Adjust the Calibrator using the output adjustment
4. Verify that the result is within the test limits.
5. Repeat the previous steps for the 2 mV output on all remaining frequencies on the
6. Proceed to the remaining output levels and frequencies list in Table 3-23 or 3-24 and
3-30
Figure 3-9.
Set the digital filter mode to FAST and the restart to MEDIUM. Push the DONE soft
key twice to return to the measurement display.
knob for a reading of 2.0000 mV ±1 count on the 5790A. Record the Calibrator error
display reading in the 90 Day column in Table 3-23 (5720A) or 3-24 (5700A) as
appropriate for the verification interval.
test record.
repeat steps 3 through 5 using the appropriate output level in each step, and the
adjustment tolerance in Table 3-8.
12 Inch
Coax Cable
Figure 3-9. AC Voltage Test Setup
Note
BNC to
Banana Adapter
5790A Side View
HI
Input 2
LO
BNC to
Banana Adapter
Calibrator Side View
HI
Output
LO
F3-6.EPS