Fluke 5700A Series II Service Manual page 136

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II Calibrator
Service Manual
Output
Level
2.1V
1V
300 mV
100 mV
30 mV
10 mV
3 mV
1 mV
3-66
Table 3-33. Wideband Absolute Error 10 Hz to 500 kHz
1 kHz Absolute
Frequency
Error
ppm
10 Hz
30 Hz
10 kHz
120 kHz
500 kHz
10 kHz
120 kHz
500 kHz
10 kHz
120 kHz
500 kHz
10 kHz
120 kHz
500 kHz
10 kHz
120 kHz
500 kHz
10 kHz
120 kHz
500 kHz
10 kHz
120 kHz
500 kHz
10 kHz
120 kHz
500 kHz
Flatness Error
Error Sum
ppm
ppm
90-Day
Spec Limit
( ppm)
5238
2238
2238
2238
2238
2900
2900
2900
2833
2833
2833
3400
3400
3400
3333
3333
3333
3900
3900
3900
4833
4833
4833
5400
5400
5400

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series ii5720a series ii

Table of Contents