5700A/5720A Series II Calibrator
Service Manual
5700A/5720A Series
II Output
2V
20V
200V
300V(1)
300V(1)
300V(1)
(1) The 5700A/5720A Series II maximum volt-Hertz product is (2.2 x 10
Fluke 5725A Amplifier is attached.
Frequency
3-68
Table 3-37. AC V Distortion Test Summary
Load Resistors
100 , 1/8W
1 k , 1/2W
10 k , 5W
15 k , 5W
15 k , 5W
15 k , 5W
Table 3-38. Test Record for Flatness Check of the AC 2 mV Range
10 Hz
20 Hz
40 Hz
1 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
1 MHz
Frequency
10 Hz, 20 Hz
1 kHz, 20 kHz, 50 kHz,
100 kHz, 200 kHz, 500 kHz
10 Hz, 20 Hz
20 kHz, 100 kHz
200 kHz, 500 kHz
10 Hz, 20 Hz
50 kHz, 100 kHz
40 Hz
50 kHz
70 kHz
UUT Error Display Reading
Measured
Distortion
Distortion
0.054%
0.044%
0.355%
0.0535%
0.0385%
0.304%
0.055%
0.1065%
0.1%
0.3%
0.4%
7
) The 300V level assumes that a
Limits
0.26%
0.26%
0.29%
0.49%
0.87%
1.83%
Max.