Table 3-30. Wideband Frequency Accuracy Test Record
Frequency ( Hz )
Frequency Measured
10 Hz
100 Hz
300 Hz
500 Hz
800 Hz
900 Hz
1 kHz
1.19 kHz
2.2 MHz
3.5 MHz
3.8 MHz
10 MHz
20 MHz
30 MHz
Determining Test Limits for Other Calibration Intervals
Tolerance Limits
99.99 ms to 100.01 ms
9.999 ms to 10.001 ms
299.97 Hz to 300.03 Hz
499.95 Hz to 500.05 Hz
799.92 Hz to 800.08 Hz
899.91 Hz to 900.09 Hz
999.0 Hz to 1.0001 kHz
1.189881 kHz to 1.190119 kHz
2.19978 MHz to 2.20022 MHz
3.49965 MHz to 3.50035 MHz
3.79962 MHz to 3.80038 MHz
9.990 MHz to 10.001 MHz
19.998 MHz to 20.002 MHz
29.997 MHz to 30.003 MHz
Calibration and Verification
3
3-63