Fluke 5700A Series II Service Manual page 135

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

Output
Frequency
Level
30 mV
10 kHz
120 kHz
500 kHz
2 MHz
5 MHz
10 MHz
20 MHz
30 MHz
10 mV
10 kHz
120 kHz
500 kHz
2 MHz
5 MHz
10 MHz
20 MHz
30 MHz
3 mV
10 kHz
120 kHz
500 kHz
2 MHz
5 MHz
10 MHZ
20 MHz
30 MHz
1 mV
10 kHz
120 kHz
500 kHz
2 MHz
5 MHZ
10 MHz
20 MHz
30 MHz
Table 3-32. Wideband Flatness Test Record (cont)
Measured
Spec
Flatness Error
ppm
(90-day)
1000
1000
1100
1100
2100
2100
4100
10100
1000
1000
1300
1300
2300
2300
4300
10300
1000
1000
2000
2000
4000
4000
6000
16000
1000
1000
5000
5000
7000
7000
9000
30000
Determining Test Limits for Other Calibration Intervals
5790A
TUR
1-Year
Spec
500
2.00
500
2.00
500
2.20
500
2.20
1000
2.10
1000
2.10
1500
2.73
3500
2.89
500
2.00
500
2.00
700
1.86
700
1.86
1000
2.30
1000
2.30
1700
2.53
3700
2.78
500
2.00
500
2.00
1033
1.94
1033
1.94
1333
3.00
1333
3.00
2033
2.95
4033
3.97
500
2.00
500
2.00
1700
2.94
1700
2.94
2700
2.59
2700
2.59
4000
2.25
9000
3.33
Calibration and Verification
Test Limit
GB
ppm
Factor
0.932
932
0.932
932
0.945
1040
0.945
1040
0.939
1972
0.939
1972
0.972
3985
0.977
9868
0.932
932
0.932
932
0.920
1196
0.920
1196
0.950
2185
0.950
2185
0.962
4137
0.972
10012
0.932
932
0.932
932
0.928
1856
0.928
1856
0.980
3920
0.980
3920
0.978
5868
0.999
15984
0.932
932
0.932
932
0.978
4890
0.978
4890
0.965
6755
0.965
6755
0.948
8532
0.988
29640
3
3-65