Table 3-24. 5700A AC Voltage Output Test Record (cont)
Output
Frequency
Level
200V
1 kHz
200V
20 kHz
200V
50 kHz
200V
100 kHz
200V(2)
40 Hz
200V
20 Hz
200V
10 Hz
300V(3)
20 kHz
600V(3)
50 kHz
600V(3)
100 kHz
1 kV
1 kHz
1 kV
50 Hz
1 kV
300 Hz
1 kV(3)
20 kHz
1 kV(3)
30 kHz
1 kV(3)
40 Hz
1: This is a test of the bottom of the 20V range .
2: Observe the Calibrator output for 10 minutes and verify that it remains stable within 7.5 ppm.
3: Perform only for units that are used with a 5725A Amplifier.
4: When the TUR (Test Uncertainty Ratio) is less than 4:1, the spec is reduced to give the same
Consumer Risk as a 4:1 TUR as described later in this chapter.
Characterized 1.9 mV Reading
Error
Spec
Display
( ppm )
Reading
85
85
240
600
85
185
600
145
378
1375
84
84
84
131
371
84
Table 3-25. AC Voltage 2 mV Range Test Record
Error Display Reading
Determining Test Limits for Other Calibration Intervals
5790A
1-Year
TUR
Spec
Factor
( ppm)
31
2.74
0.971
31
2.74
0.971
69
3.48
0.991
98
6.12
1.00
31
2.74
0.972
68
2.72
0.970
200
3.00
0.980
41
3.54
0.993
130
2.91
0.977
500
2.21
0.945
38
2.21
0.945
38
2.21
0.945
38
2.21
0.945
38
3.45
0.991
130
2.85
0.975
38
2.21
0.945
Calibration and Verification
GB
Test Limit (4)
( ppm)
83
83
238
83
179
588
144
369
1299
79
79
79
130
362
79
Limits
0.26%
3
3-59