Fluke 5700A Series II Service Manual page 123

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

Divider
5700A
Setting
Range
0.1V
0.22V
1V
2.2V
1V
11V
10V
11V
10V
22V
100V
220V
1000V
1100V
0.1V
0.22V
1V
2.2V
10V
11V
1: Mathematically, the true 5700A/5720A Series II output programmed is the certified value of the
reference standard divided by the reference standard nominal value, multiplied by the required
5700A/5720A Series II nominal output. In other words, the 5700A/5720A Series II output is always
programmed for the nominal output adjusted up or down by the same percentage as the certified value
of the reference standard.
2: On the 752A 0.1 and 1V ranges, the null detector polarity is reversed. A low input (5700A/5720A
Series II output) causes a positive null detector reading.
3: Use Range Lock to obtain 1V on 11V and 10V on 22V range. Deactivate Range Lock before setting
the next voltage output.
4: Line regulation can be verified at this time by adjusting the autotransformer for a 10% change in line
voltage. The null detector reading must remain constant within 1 V.
Table 3-19. DC Voltage Test Record for 5700A
5700A
Null Detector
Output
Reading( V)
(1)
0.1V
1V
1V (3)
10V (4)
10V (3)
100V
1000V
-0.1V
-1V
-10V
Determining Test Limits for Other Calibration Intervals
Null Det Limit
( V)
(2)
1.45 (2)
7.2 (2)
9 (2)
54
58
70
86
1.45 (2)
7.2 (2)
54
Calibration and Verification
Meter Limit
Rattle ( V)
NA
0.55
2.2
3.5
5.5
7.5
4.5
NA
0.55
3.5
3
3-53