When verifying the 5725A Amplifier at the 10A level, allow sufficient time
for the A40A-10 shunt to reach thermal stability after initially applying the
current.
A40 Current Shunt
5790A
Input 1
792A-7004
A40 Current Shunt Adapter
A. Alternating Current Test Setup, 22 mA to 2A Using 5790A Input 1
Red
5790A
Input 1
Black
792A-7004
A40 Current
Shunt Adapter
B. Alternating Current Test Setup, 10A Using 5790A Input 1
Figure 3-10. Alternating Current Test Setup
Note
Red
Pomona Cable
1368-A-18
Black
Pomona Cable
5268-C-12
UHF(M) to BNC(F)
Adapter
Kings KC-99-34
Calibration and Verification
Calibrator (UUT)
Hi
Lo
A40A 10A
Current Shunt
FLUKE A45-4003
P/N 212852
Cable
Pomona Model 1707
UHF(F) to Banana
Adapter
Full Verification
Calibrator (UUT)
5725A (UUT)
Hi
Lo
F3-8A.EPS