Calibrating The Wideband Ac Module (Option 5700A-03) - Fluke 5700A Series II Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

HI
LO
GUARD
752A
INPUT
NULL
DETECTOR
REFERENCE
STANDARD

3-8. Calibrating the Wideband AC Module (Option 5700A-03)

The Wideband AC Module (Option 5700A-03) can be installed in either the 5720A or
5700A Series II calibrator. The module needs to be calibrated for both gain and flatness.
The gain should be calibrated when the 5700A-03 main output functions undergo their
routine calibration.
NULL DETECTOR
GND
NC
VOLTAGE DIVIDER STANDARD
LO
HI
LO
HI
OUTPUT
GND
HI
LO
HI
LO
Figure 3-3. 220V DC Range Calibration Connections

Calibrating the Wideband AC Module (Option 5700A-03)

EX SNS
GRD
732B
732B DC STANDARD
AC PWR
10V
10V COM
SERIAL NUMBER
Calibration and Verification
: OFF
: OFF
EX GRD
CALIBRATOR
OUTPUT
SENSE
V A
VA
WIDEBAND
HI
HI
LO
LO
HI
AUX
GUARD GROUND
CURRENT
VOLTAGE REFERENCE
STANDARD
IN CAL
CHARGE
LOW BAT
1.018V
CHASSIS
1.018V COM
GUARD
F7-6.EPS
3
3-15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series ii5720a series ii

Table of Contents