3-8. Calibrating the Wideband AC Module (Option 5700A-03)
The Wideband AC Module (Option 5700A-03) can be installed in either the 5720A or
5700A Series II calibrator. The module needs to be calibrated for both gain and flatness.
The gain should be calibrated when the 5700A-03 main output functions undergo their
routine calibration.