Table 3-27. 5700A AC Current 20 mA to 10A Accuracy Test Record
2A, 200 mA, 20 mA and 10A Test Record (10A Only for 5725A)
Output
Freq
Current
2A
40 Hz
2A
1 kHz
2A
5 kHz
2A
10 kHz
200 mA
10 Hz
200 mA
20 Hz
200 mA
40 Hz
200 mA
1 kHz
200 mA
5 kHz
200 mA
10 kHz
20 mA
10 Hz
20 mA
20 Hz
20 mA
40 Hz
20 mA
1 kHz
20 mA
5 kHz
20 mA
10 kHz
10A (1)
40 Hz
10A (1)
1 kHz
10A (1)
5 kHz
10A (1)
10 kHz
1: If a 5725 is attached.
Determining Test Limits for Other Calibration Intervals
A40 Shunt
UUT Error
AC-DC Diff
Display
Calibration and Verification
Calculated
Specs
Error
ppm
670
670
800
9100
725
400
170
170
850
2100
725
400
160
60
850
2100
417
417
888
3375
3
3-61