Guidance and manufacturer's declaration-electromagnetic immunity
for all EQUIPMENT AND SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration-electromagnetic immunity
The
19CX/22CX Touchcomputer for Healthcare Applications
the electromagnetic environment specified below. The user of the
Touchcomputer for Healthcare Applications should assure that it is used in such an
Immunity
IEC 60601 Test
Test
Conducted RF
3 Vrms
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5
GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Compliance
Level
Level
3 Vrms
3 Vrms
C-Series Touchcomputer for Healthcare Applications xi
is intended for use in
19CX/22CX
Electromagnetic
Environment-Guidelines
Portable and mobile RF
communications equipment should be
used no closer to any part of the
19CX/22CX Touchcomputer for
Healthcare Applications and should
assure that it is used in such an
environment, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
80MHz to 800 MHz
800 NHz to 2.5GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance
in metres(m)
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range
Interference may occur in the
vicinity of equipment marked with
the following symbol:
1
, should
2
.