Agilent Technologies InfiniiVision 2000 X-Series User Manual page 122

Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

9
Triggers
5 Press the pulse polarity softkey to select positive (
6 Press the qualifier softkey (< > ><) to select the time qualifier.
122
For digital channels, press the [Digital] key and select Thresholds to set
the threshold level.
The value of the trigger level or digital threshold is displayed in the
upper- right corner of the display.
polarity for the pulse width you want to capture.
The selected pulse polarity is displayed in the upper- right corner of the
display. A positive pulse is higher than the current trigger level or
threshold and a negative pulse is lower than the current trigger level or
threshold.
When triggering on a positive pulse, the trigger will occur on the high
to low transition of the pulse if the qualifying condition is true. When
triggering on a negative pulse, the trigger will occur on the low to high
transition of the pulse if the qualifying condition is true.
The Qualifier softkey can set the oscilloscope to trigger on a pulse
width that is:
Less than a time value (<).
For example, for a positive pulse, if you set t<10 ns:
10 ns
Greater than a time value (>).
For example, for a positive pulse, if you set t>10 ns:
10 ns
Within a range of time values (><).
For example, for a positive pulse, if you set t>10 ns and t<15 ns:
10 ns
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
) or negative (
10 ns
Trigger
Trigger
)