Trigger Indicator
The trigger indicator at the top right of the display shows whether triggers
are occurring.
In the Auto trigger mode, the trigger indicator can show:
• Auto? (flashing) — the trigger condition is not found (after the
• Auto (not flashing) — the trigger condition is found (or the pre- trigger
In the Normal trigger mode, the trigger indicator can show:
• Trig'd? (flashing) — the trigger condition is not found (after the
• Trig'd (not flashing) — trigger condition is found (or pre- trigger buffer is
When the oscilloscope is not running, the trigger indicator area shows
Stop.
When to Use Auto
The Auto trigger mode is appropriate when:
Trigger Mode
• Checking DC signals or signals with unknown levels or activity.
• When trigger conditions occur often enough that forced triggers are
When to Use
The Normal trigger mode is appropriate when:
Normal Trigger
• You only want to acquire specific events specified by the trigger
Mode
• Triggering on an infrequent signal from a serial bus (for example, I2C,
• Making single- shot acquisitions with the [Single] key.
See Also
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
pre- trigger buffer has filled), and forced triggers and acquisitions are
occurring.
buffer is being filled).
pre- trigger buffer has filled), and no acquisitions are occurring.
being filled).
unnecessary.
settings.
SPI, CAN, LIN, etc.) or another signal that arrives in bursts. The Normal
trigger mode lets you stabilize the display by preventing the oscilloscope
from auto- triggering.
Often with single- shot acquisitions, you must initiate some action in the
device under test, and you don't want the oscilloscope to auto- trigger
before that happens. Before initiating the action in the circuit, wait for
the trigger condition indicator Trig'd? to flash (this tells you the
pre- trigger buffer is filled).
"Forcing a Trigger"
"To set the trigger holdoff"