Acquisition Control
Glitch or Narrow Pulse Capture
A glitch is a rapid change in the waveform that is usually narrow as
compared to the waveform. Peak detect mode can be used to more easily
view glitches or narrow pulses. In peak detect mode, narrow glitches and
sharp edges are displayed more brightly than when in Normal acquire
mode, making them easier to see.
To characterize the glitch, use the cursors or the automatic measurement
capabilities of the oscilloscope.
Figure 24
Sine With Glitch, Normal Mode
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide