6712FR Refrigerated Sampler
Section 5 Extended Programming
Table 5-2 Storm Water Runoff Sampling (Continued)
PACED BY:
FLOW PULSES
48
FLOW MODULE VOLUME
FLOW BETWEEN
SAMPLE EVENTS:
49
2500 gal
(0.001-99999)
SAMPLE AT START?
50
YES NO
1 BOTTLES PER
51
SAMPLE EVENT (1- 18)
SWITCH BOTTLES ON:
NUMBER OF SAMPLES
52
TIME
SWITCH BOTTLES EVERY
53
1 SAMPLES (1- 50)
RUN CONTINUOUSLY?
54
YES
NO
SAMPLE VOLUME:
55
200 ml (10-1000)
ENABLE:
RAIN LEVEL FLOW
56
'A' DONE NONE
ENABLE: RAIN
57
AND
OR
DONE
ENABLE: RAIN AND
LEVEL FLOW 'A' DONE
58
DONE
"RAIN" SET POINT:
0.15 INCHES PER
59
__HOURS, __MINUTES
ENABLED WHEN:
ABOVE SET POINT
60
BELOW SET POINT
5-10
Select FLOW MODULE VOLUME.
Enter 2500.
Select YES. This will cause the first sample to be taken 20 minutes (the
delay to start of sampling specified in Step 67) after the enable for Part B.
Enter 1.
Select NUMBER OF SAMPLES.
Enter 1.
Select NO.
Enter 500.
Select RAIN.
Select AND.
Select LEVEL.
Enter 0.15 for inches, then enter 30 minutes.
Select ABOVE SET POINT.