8.4.3 Internal Battery
Warning
Table 8-3 Checking and Replacing the Internal Battery
RUN
PROGRAM
1
VIEW REPORT
OTHER FUNCTIONS
MAINTENANCE
MANUAL FUNCTIONS
2
PROGRAMMING STYLE
HARDWARE
SET CLOCK
PUMP TUBE ALARM
3
INTERNAL BATTERY
DIAGNOSTICS
WARNING!
REPLACE INTERNAL
4
BATTERY BEFORE
17-OCT-97
8-6
Table 8-2 Resetting the Pump Counter
Select OTHER FUNCTIONS.
Select MAINTENANCE.
Select PUMP TUBE ALARM.
The sampler displays this screen briefly to give you the current pump-count
information. The first line shows you the pump counts tallied since the
counter was last reset. The second line shows the current alarm setting.
To reset the counter to zero, select YES. Always reset the counter after
replacing a pump tube. Select NO when merely checking the current count.
If necessary, change the pump-count alarm setting by typing the first two
digits of the new setting. The sampler accepts entries between 1 and 99. For
example, to increase the count to 1,500,000, enter 15.
STOP
Press
to return to the main menu.
The internal lithium battery preserves the sampler's memory
when power is removed. For an explanation of how to replace the
lithium battery, go to Section 8.8
.
Select OTHER FUNCTIONS.
Select MAINTENANCE.
Select INTERNAL BATTERY.
The sampler displays this screen briefly to show you the expiration date.