Figure 5-4 Start Time Diagram for Sampler Enable Settings
When the sampler detects that it is disabled at start time, it sus-
pends the program until it is enabled. Once enabled, the sampler
collects a sample only when programmed to take a sample at the
enable time. The diagram in Figure 5-4 shows the sampler's
response when enabled or disabled at the start time.
Sampler enabled at
start time
A sample event at the start time
always occurs for time-paced programs
(except for random). For flow-paced
programs, it occurs when the program
requires a sample at start time.
Sampler disabled at
start time by
signal from pin F
A sample event at enable time occurs
only when the program requires a
sample at enable.
6712FR Refrigerated Sampler
Section 5 Extended Programming
5-23