6712FR Refrigerated Sampler
Section 1 Introduction
Table 1-3 Technical Specifications for the 6712FR Refrigerated Sampler (Continued)
Dry Weight:
Dry Weight With:
Filled Bottle Weight:
Average Thermal
Resistance Factor:
Recovery time
Recovery time is the time for the internal air temperature to return to 39°F (4°C) after the door has been opened for
one minute. The empty refrigerator is running and set at 39°F (4°C). The door is opened at least 4 minutes after the
compressor has turned on. Ambient air temperature is 75°F (24°C).
Thermostat
Temperature Ranges
Environment
Electrical
1-12
145 lbs (66 kg) Refrigerator only, no controller or bottle configuration
160 lbs (73 kg) Includes refrigerator, controller, top cover, center section, tub,
pump tube, distributor shaft extension, distributor arm and nut, and discharge
tube for the 24 plastic bottle configuration.
24 plastic bottles, rack, hold down, no caps
24 glass bottles, rack, expander ring, no caps
12 plastic bottles, rack, hold down, no caps
8 plastic bottles, rack, adaptor, no caps
8 glass bottles, rack, adaptor, no caps
2 plastic bottles, rack, no caps
2 glass bottles, rack, no caps
1-9.4 l plastic bottle and cap, no distributor arm or nut165 lbs(75 kg)
1-9.4 l glass bottle and cap, no distributor arm or nut 170 lbs(77 kg)
1-15.0 l plastic bottle and cap, no distributor arm or nut165 lbs(75
kg)
1-20.5 l plastic bottle and cap, no distributor arm or nut165 lbs(75
kg)
1-18.5 l glass bottle and cap, no distributor arm or nut 170 lbs(77 kg)
24 plastic bottles
101 lbs(46 kg)
24 glass bottles
104 lbs(48 kg)
12 plastic bottles
89 lbs (41 kg)
12 glass bottles
95 lbs (43 kg)
8 plastic bottles
84 lbs (38 kg)
8 glass bottles
90 lbs (41 kg)
4 plastic bottles
98 lbs (45 kg)
4 glass bottles
104 lbs(48 kg)
1 plastic bottle
90 lbs (41 kg)
1 glass bottle
94 lbs (43 kg)
R-16
The average R-value for a complete FR refrigerator not running
5 minutes
Set point range: 32° to 46°F (0° to 8°C)
Set point accuracy: ±1.8°F (±1°C) at 39°F (4°C)
[With an ambient air temperature of 75°F (24°C)]
Operational: -20° to 120°F (-29° to 49°C)
Storage: 0° to 140°F (-18° to 60°C)
Rain and spray resistant, but not watertight. Do not install in locations where the
refrigerator's lower compartment could become submerged.
Power Requirements: 120 Volts, 60 Hz
Running current:
Starting current:
Stalled Compressor current:21 amp
(maximum)
3.5 amp
17 amp
175 lbs(79 kg)
195 lbs(88 kg)
175 lbs(79 kg)
170 lbs(77 kg)
180 lbs(82 kg)
175 lbs(79 kg)
165 lbs(75 kg)
230 Volts, 50 Hz
(optional)
1.75 amp
8.5 amp
10.5 amp