Teledyne 6712FR Installation And Operation Manual page 60

Refrigerated sampler
Table of Contents

Advertisement

6712FR Refrigerated Sampler
Section 4 Standard Programming
Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample (Contin-
NUMBER OF BOTTLES:
4
1 2 4 8 12 24
BOTTLE VOLUME IS
5
1000 ml (300-30000)
SUCTION LINE LENGTH
6
IS 10 ft
(3-99)
TIME PACED
7
FLOW PACED
FLOW BETWEEN
SAMPLE EVENTS:
8
2 PULSES (1-9999)
SEQUENTIAL
BOTTLES/SAMPLE
9
SAMPLES/BOTTLE
RUN CONTINUOUSLY?
10
YES NO
1 BOTTLES PER
11
SAMPLE EVENT (1-24)
NO DELAY TO START
DELAYED START
12
CLOCK TIME
WAIT FOR PHONE CALL
START FLOW COUNT AT:
00:00
13
SELECT DAYS:
SU MO TU WE TH FR SA
14
DONE
PROGRAMMING COMPLETE
RUN THIS PROGRAM
15
NOW?
YES NO
4-4
Select 24.
Enter 1000.
Enter 7.
Select FLOW PACED.
Enter the number of pulses between sample events.
Select BOTTLES/SAMPLE.
For this example, select NO by pressing an arrow until the option blinks.
Then, press
. Selecting YES allows the program to run indefinitely by
repeating the sample distribution. Continuous sampling assumes that filled
bottles are replaced with empty bottles at regular service intervals.
Enter 2.
Select CLOCK TIME.
Enter 6, then enter 0.
This is the time at which the sampler will start the pacing countdown.
Select MO, WE, and FR.
Use the arrow keys and
blinking, select DONE and press
Select NO.
to select the days. When the correct days are
↵.