6712FR Refrigerated Sampler
Section 5 Extended Programming
ENABLED WHEN:
ABOVE SET POINT
5
BELOW SET POINT
EVENT02:
RAIN LEVEL FLOW
6
TEMP5
pH5
SP_CO5
DO5 DONE
"LEVEL" CONDITION:
SET POINT
7
RANGE
RATE OF CHANGE
CONDITION IS TRUE
WHEN "LEVEL"
8
RISES
FALLS
"LEVEL" RISES
1.000 ft
9
__HOURS, __MINUTES
EVENT03:
RAIN LEVEL FLOW
10
TEMP5
pH5
SP_CO5
DO5 DONE
"pH5" CONDITION:
SET POINT
11
RANGE
RATE OF CHANGE
"pHx" RANGE:
LOWER:
4.00
12
UPPER: 10.00
( 0.00 - 14.00)
ENABLED WHEN:
INSIDE RANGE
13
OUTSIDE RANGE
EVENT04:
RAIN LEVEL FLOW
14
TEMP5
pH5
SP_CO5
DO5 DONE
PACING:
EVENT PACED
15
3 EVENT CONDITIONS
5-14
Table 5-3 Event Paced Sampling (Continued)
Select ABOVE SET POINT.
Select LEVEL again.
Choose RATE OF CHANGE.
Choose RISES.
This will enable the event condition if a rise occurs within the time frame
and above the amount specified in screen #9.
Enter 1 ft and press
Then enter 2 hours.
Select pH.
Select RANGE.
Enter 4 for the lower range and 10 for the upper range.
Choose INSIDE RANGE.
You will not be entering a fourth event condition, so use the arrow keys to
scroll down, and then select DONE.
Proceed with the rest of the programming steps.
.