Replacing The Internal Battery; Main Circuit Board Connections - Teledyne 6712FR Installation And Operation Manual

Refrigerated sampler
Table of Contents

Advertisement

6712FR Refrigerated Sampler
Section 8 Maintenance
Keypad Connector
Battery
Liquid Detector
Connector
Figure 8-4 Main Circuit Board Connections
8.8 Replacing the Internal
Battery
Battery Status Display
WARNING!
REPLACE INTERNAL
BATTERY BEFORE
17-OCT-06
DID YOU REPLACE THE
INTERNAL BATTERY?
YES
8-14
Connector Board
Connector
A lithium battery housed inside the controller maintains power
to the sampler's memory when the controller is disconnected
from a power source. If it discharges completely, the sampler will
lose all program settings and all data stored in memory when the
external power source is disconnected.
The battery requires replacement every five years. The sampler
keeps track of the time that has passed since the battery was
installed. To view this information, select INTERNAL BATTERY
from the Maintenance screen (see example at left). After 5
seconds, the sampler asks if you have replaced the internal
battery (see example at left).
Select YES to reset the countdown to 59 months from the current
date.
While replacing the battery, it is likely that the stored readings
NO
will be lost and your program will revert to the default program.
Before proceeding, retrieve the stored data and record your
program settings.
Note
Teledyne Isco recommends you replace the battery every five
years, as soon as the battery warning appears. To ensure that
the replacement meets Teledyne Isco's specifications, use only
Isco replacement batteries. (See Appendix D, Replacement
Parts.)
Module Connector
Pump Count
Connector