Table 5-2 Storm Water Runoff Sampling (Continued)
PROGRAM MODULE?
11
YES
NO
MODE OF OPERATION
FLOWMETER
12
LEVEL ONLY
WEIR FLUME
DATA POINTS
13
EQUATION MANNING
FLOW-INSERT
WEIR/ORIFICE TYPE
V-NOTCH
14
ROUND
FLOW INSERT SIZE:
6" 8" 10" 12"
15
NEW MODULE SETUP--
DOWNLOAD DATA NOW
16
OR LOSE ALL DATA!
DONE
CURRENT LEVEL IS
___.___ ft.
17
ADJUST LEVEL TO
___.___ ft.
DATA STORAGE
INTERVAL IN MINUTES
18
1 2 5
10 15 30
NUMBER OF BOTTLES:
19
1
2
4
8
12
BOTTLE VOLUME IS
20
1000 ml (300-30000)
SUCTION LINE LENGTH
21
IS 5 ft
(3-99)
AUTO SUCTION HEAD
22
ENTER HEAD
Select YES
Select FLOWMETER
Select FLOW-INSERT.
Select ROUND.
Select 12.
Select DONE.
Note: Partition memory status message will be displayed briefly.
Press ENTER.
Select 5. (If the data storage interval is different from the current program,
you will be informed that the storage interval has changed, and that the
stored data will be lost. Download stored data (if desired) and select DONE.
Select 24.
24
Enter 1000.
Enter 5. The 6712FR calculates new pump tables.
Select AUTO SUCTION HEAD.
6712FR Refrigerated Sampler
Section 5 Extended Programming