Chemical/Dwell

The Chemical/Dwell program should be used when disinfecting the hydraulic system using
chemical disinfectants that can remain in contact with internal components for prolonged
periods without damaging them. Diacide HD or formaldehyde can be used with the
Chemical/Dwell program for maximum effectiveness.
To run the Chemical/Dwell program:
1. Attach a sign to the front of the machine that identifies the chemical being used to
disinfect the machine.
2. Ensure that both dialysate lines are on the shunt.
3. Place the concentrated disinfectant in the small container with the yellow cap.
4. Place both concentrate connectors in their respective ports. The "Select Program" screen
appears in the control panel display.
5. From the "Select Program" screen, touch Chemical/Dwell.
Figure 53 – The Chemical/Dwell Screen
6. The "Chemical/Dwell" screen appears in the display (see Figure 53). The progress of the
disinfection program is indicated by the horizontal bar. The program starts with a 45
second rinse. The message, "Rinsing Line, Please Wait" is displayed in the Status box.
7. When prompted, connect red connector to chemical disinfectant. Press CONFIRM.