Message
Purpose of Message
Fail* Venous High Soft
Venous Pressure test has failed.
Fail* Venous Low Soft
Fail* Venous High Hard
Fail* Venous Low Hard
Filling Program Alarm
A Fill program has occurred for one
minute while blood is sensed.
Fill for Diasafe test
This message indicates the status of
the Diasafe test
Filling Program
A Fill program is in progress.
Flow Error
General Flow Alarm
147
Type
Action Required
Test
Rerun test, if failure repeats, remove from service.
Message
Alarm
Inspect for air in the system.
Correct as required.
If the warning is not cleared, take the machine out of service.
Advisory
Advisory only. No action is required.
Advisory
Advisory only. No action is required.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly:
1) Check the water supply flow to the machine.
2) Check that the Dialysate Flow is on.
3) Check the dialysate lines for kinks.
4) Set Dialysate Flow in the "Home" screen to 500 ml/min and verify
that the flow from the drain line is 500 ml/min  50 ml/min.
5) Turn the power off and on. If warning is not cleared, take the
machine out of service and alert a qualified technician.
Troubleshooting
2008K Operator's Manual rev. K