Fresenius Medical Care 2008K Operator's Manual page 197

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Cuff Pressure Range
Initial Cuff Inflation
Systolic Determination Range
MAP Determination Range
Diastolic Determination Range
Pulse Rate Determination Range
Cuff Inflation Rate
Determination Time Normal
Overpressure Cut Off
Transducer Drift
Leakage Rate (Max)
Pressure Rate Offset
Alarm Preset Values (Internal alarm values
preset to provide alarm limits in the event
individual values are not entered)
Systolic
MAP
Diastolic
Pulse
Inflation Pressure
2008K Operator's Manual rev. K
Performance Limits
Pediatric
5-220 mmHg
125 mmHg or adjusted by
host
40-160 mmHg
25-140 mmHg
15-120 mmHg
40-240 BPM
5 seconds
Approx. 20 seconds
220 mmHg
Auto Zeroing
3 mmHg/min in 3 minutes
Auto Zeroing
Pediatric/Neonatal
160/80
120/70
100/60
120/50
Auto
Machine Specifications
Adult
10-325 mmHg
180 mmHg or adjusted by
host
80-260 mmHg
35-220 mmHg
25-200 mmHg
40-180 BPM
5 seconds
25-30 seconds
325 mmHg
Auto Zeroing
3 mmHg/min in 3 minutes
Auto Zeroing
Adult
200/90
120/70
110/50
120/50
Auto
197