11. When the air has been cleared from the heparin line, release the CONFIRM key. The pump will
stop.
Warning! Clamp the heparin line closest to the "T" connection during recirculation if
using reuse dialyzer.
12. In the "Heparin" screen, set the treatment parameters for Rate, Infusion Time, and Bolus as
described in Figure 34 – Heparin Delivery Screen on page 72.
The heparin administration system is now ready for patient treatment.
Warning! The heparin pump is to be used only under positive pressure conditions.
Under negative pressure conditions, excessive heparin may be infused.