Fresenius Medical Care 2008K Operator's Manual page 29

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Alarm Light (red)
Start/Stop Key
Arterial Pressure
Port (P
)
Art.
Level Adjust Key
Tubing Retainer
Figure 9 – The Blood Pump Module
The following table describes the operational features of the blood pump.
Table 6- Blood Pump Features
FEATURE
Start/Stop Key
Pressure Port
Level Adjust Key
Display Window
Up/Down Keys
Tubing Retainer
2008K Operator's Manual rev. K
PURPOSE
Starts and stops the blood pump. Opening the door will also stop the
blood pump.
Line from arterial drip chamber is connected to a transducer protector
and attached here to provide arterial blood pressure readings.
Pressing the
key (level adjust key on the Blood Pump module) will
raise the level of fluid in the arterial drip chamber.
Displays the blood flow rate setting in increments of 5 ml/min during
blood pump operation. When the door is open it displays the pump-
segment diameter in mm.
Increases the speed of the pump when Up arrow (
decreases the pump speed when Down arrow (
door is open, simultaneously press the
the
or
key to select the pump segment diameter.
A spring-loaded device that secures the pump segment in place.
Display Window
Up/Down Keys
Rotor
Yoke
) is depressed,
) is depressed. When
and
keys and then press
Modules
29