The battery RAM memory has been
lost.
Lower Dia. has been set
The operator has attempted to set the
to Min [Max]
lower diastolic pressure limit higher or
lower than allowed.
Lower Pulse has been set
The operator has attempted to set the
to Min [Max]
lower pulse rate limit higher or lower
than allowed.
Lower Sys. has been set
The operator has attempted to set the
to Min [Max]
lower systolic pressure limit higher or
lower than allowed.
Max UF rate reached.
This message informs the operator
Select new Goal or Time
that the calculated UF rate is higher
than the internal selection allows.
Max UF time reached
This message informs the operator
that the calculated UF time is higher
than the maximum allowed.
153
Type
Action Required
Opening
Verify all treatment settings before using the machine.
Screen
Message
Dialog
The machine has set the limit to the highest or lowest value allowed.
Message
Verify that the limit setting is acceptable
Dialog
The machine has set the limit to the highest or lowest value allowed.
Message
Verify that the limit setting is acceptable
Dialog
The machine has set the limit to the highest or lowest value allowed.
Message
Verify that the limit setting is acceptable
Dialog
In the "Home" screen, decrease the UF Goal or increase the UF
Message
Time.
Dialog
In the "Home" screen, decrease the UF Time
Message
Troubleshooting
2008K Operator's Manual rev. K