Fresenius Medical Care 2008K Operator's Manual page 131

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Message
Purpose of Message
ACID PUMP ALARM
This is a pump failure warning.
ACID PUMP ALWAYS
This is a pump failure warning.
EOS
ACID PUMP NO EOS
This is a pump failure warning.
Act Blood Pump Failed
Electronic self-test failure.
Act Board CRC Error
Electronic self-test failure.
Act BYP Valve Fail 1
Electronic self-test failure.
Act BYP Valve Fail 2
Electronic self-test failure.
Actuator BD No Echo
Functional to Actuator board
communication problem
Adjusting TMP
The operator has chosen to relieve
the TMP after a TMP alarm
131
Type
Action Required
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn power off and back on. If warning is not cleared, take
the machine out of service.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn power off and back on. If warning is not cleared, take
the machine out of service.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn power off and back on. If warning is not cleared, take
the machine out of service.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Advisory
No action necessary.
Troubleshooting