Fresenius Medical Care 2008K Operator's Manual page 141

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Message
Purpose of Message
Conductivity Low,
continued
Conductivity Limits Set to
Conductivity limits were found to be
Default
outside of the allowed range at power
up.
Conductivity out of range
When attempting to start an OLC test,
certain conditions are necessary,
including the condition that
conductivity for both the inlet and
outlet sensors be in range
Confirm Concentrate
This message will be displayed if the
user needs to confirm the concentrate
selected for use
Connector(s) Out Of Port
An action has been initiated that
requires the Acid/Bicarbonate
Connectors to be in their rinse ports
Cooling Down
The machine is cooling down from a
heat disinfect.
Cover is Open
The dialysate shunt door is open.
Cuff Pressure = XXX
This is displayed during the blood
pressure measurement. The cuff
pressure is XXX mmHg.
141
Type
Action Required
4) Turn off the dialysate flow. Reconnect the connector to the wand
and turn on the dialysate flow. Allow 5 minutes for the
conductivity to reach the prescribed level. Verify correct
conductivity and pH before beginning treatment with a patient.
Sample the dialysate fluid from the blue dialysate connector and
verify the conductivity and the pH using independent testing
devices.
If conductivity alarm persists, discontinue treatment and remove
patient from the machine. Perform an acid cleaning followed by a
complete rinse cycle. Test machine operation. If conductivity alarm
persists, take the machine out of service and alert a qualified
technician
Opening
After entering Dialysis, verify that the conductivity limits are as
Screen
desired.
Advisory
Wait until the conductivity is stable and start the OLC test again. If
the message repeats, do not use OLC until the conductivity sensors
have been recalibrated.
Advisory
Press CONFIRM or change the concentrate selection and then press
CONFIRM.
Warning
Insert the concentrate connectors into their proper rinse ports.
Advisory
Advisory message only. No action is required.
Advisory
To proceed with the selected operation, close the shunt door.
Dialog
No action is necessary.
Message
Troubleshooting
2008K Operator's Manual rev. K