What to do from this screen...
Enter the prescribed dialysate settings for:
Concentrate type
Base Na
Bicarbonate level
Sodium Variation (SVS) profile
Set Alarm limits for:
Low Acid/Bicarbonate warning
Position and width of Conductivity Alarm window
Warning! The specific concentrate and sodium and bicarbonate settings must be prescribed
by a physician.
Note: If the machine is set up for use with Citrasate® or DRYalysate®, a 'Citrate' meter box
will be displayed in the dialysate constituent list.
Concentrate Type
Theoretical
Conductivity (TCD)
of dialysate at the
displayed Base Na
and Bicarbonate
values.
Electrolyte Profile of
Selected
Concentrate at the
base sodium and
bicarbonate levels
Figure 21 – The Dialysate Screen
The following table describes the features that can be programmed by the operator in the
"Dialysate" screen.
2008K Operator's Manual rev. K
+
level
Alarm Window
Conductivity
Scale
Dialysate Screen Settings
Actual Conductivity
51