molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
AUTOZ function: To periodically offset the molbox RFM RPTs relative to a reference
pressure value in order to compensate for possible changes in the RPT zero between full
recalibrations (see Section 3.4.4.4).
BPR function (present only in molbloc-S operation): To measure and display the
molbloc-S back pressure ratio (BPR) while operating in a BPR mode, which does not
continuously read and display BPR (see Section 3.4.4.5).
OPERATION
Pressing [TARE] accesses a display with the choice
of four functions (five functions with molbloc-S).
Select the desired function.
principles and operation:
Some screens (e.g., the Tare menu) may go beyond the two lines provided by the display.
This is indicated by a flashing arrow in the second line of the display. Press the [←] and
[→] keys to move the cursor to access the lines that are NOT visible or directly enter the
number of the hidden menu choice if you know it.
3.4.4.1
The purpose and operation of the tare function is different for molbloc-L and
molbloc-S operation. Tare values are generated and stored independently for
the two molbloc types. For example, a tare value generated during molbloc-L
operation is not used during molbloc-S operation, but is saved and used again
when a molbloc-L is connected to the molbox.
Tare is described separately for the two molbloc types below.
3.4.4.1.1
To zero the molbox RFM differential pressure reading at the molbloc
operating pressure.
absolute transducers as well as the microrange transducer if present.
The molbox RFM TARE function can be considered the equivalent of the
ZEROING
making measurements. molbox RFM calculates the flow through the molbloc
from the differential pressure across the molbloc.
across the molbloc is measured by taking the difference in the absolute pressure
measured by the molbox RFM's two internal RPTs. One RPT is connected to
the upstream molbloc pressure port and the other to the downstream molbloc
pressure port. If the molbox RFM has the microrange option (see Section 3.1.5),
when the differential pressure is under 12.5 kPa (1.8 psi), the microrange
option's low differential RPT is used as the source of the differential pressure
measurement.
If a common pressure is applied to both absolute transducers (and both legs of
the microrange option differential RPT if present) the differential pressure
indicated should be zero.
indicated represents an offset in the differential measurement, which will appear
as an offset or "zero error" on the flow through the molbloc calculated by the
molbox RFM. The TARE function allows the differential indication between the
two RPTs to be zeroed at the molbloc operating pressure to eliminate the zero
error in differential pressure measurement.
© 1998-2007 DH Instruments, a Fluke Company
See the following for
<1TARE>

molbloc-L OPERATION

PURPOSE
Zeros the differential between the two high pressure
PRINCIPLE
function
performed
Page 40
1tare 2purge
3leak check 4AutoZ ↓
5BPR
on
many
If a differential pressure is observed, the value
instruments
prior
to
The differential pressure