Fluke MOLBOX RFM Operation And Maintenance Manual page 135

(ver 1.10 and higher) reference flow monitor
Table of Contents

Advertisement

TARESET(=)
Purpose
To automatically tare the upstream and downstream RPT
(and microrange RPT if present) or set the tare to a user defined
value.
Syntax
"TARESET=tare(,microrange tare)
"TARESET"
Defaults
"TARESET=0(,0)"
Arguments
tare:
microrange
tare:
Remarks
"TARESET" will automatically tare the molbox RFM only if the
tare results are within the limits given (see the "TARE"
command). Use the "TARESETUP" command to prepare the unit
to be tared, and the "TARE" command to check if conditions are
valid.
After you have completed taring the unit, use the
"TARESETUP" command or the "ABORT" command to return the
unit to normal operating conditions. If the microrange option is
installed, it is also tared. You can also manually set the tare
value(s) using the "TARESET=tare" command.
values are always in Pascal [Pa].
executing, a 'b' ("busy") will appear in the third character position
of the "FR" and the "SR" command replies.
NOTE: This command typically takes 5 to 7 seconds to complete
if performing a tare (no arguments given).
Example
Command:
Reply:
Reply:
Errors
ERR# 6:
ERR# 25:
See Also
3.2.4.1, "TARE", "TARESETUP", "ABORT"
TEMP
Purpose
Read the temperature of the active molbloc in the current unit
of measure.
Syntax
"TEMP"
Remarks
The upstream and downstream temperatures as well as average
molbloc temperature are returned in the current temperature
units. They are replied in the order of:
Average, upstream, downstream
An error will be returned if there is a problem reading the
molbloc PRTs.
Example
Command:
Reply:
Errors
ERR# 37:
See Also
3.2.5, "TUNIT"
Page 125
The tare value (-9999 to 9999 Pa).
The optional microrange transducer tare value (-
999 to 999 Pa).
"TARESET"
"66 Pa"
"66 Pa, 7 Pa" (microrange option installed)
An argument is invalid.
One of the tare values exceed tare limits.
"TEMP"
"17.85 C, 17.84 C, 17.86 C "
There is an error in the molbox RFM PRT
measurement.
© 1998-2007 DH Instruments, a Fluke Company