∗OPT?
Purpose
Reads the list of installed molbloc RFM options.
Syntax
"∗OPT?"
Remarks
This query returns any registered option(s) installed in the
molbox RFM. Each option is separated by a comma.
Query
A comma delimited text field of the installed options
Reply
Example
Command:
Reply:
Reply:
∗RST
Purpose
Resets the molbox RFM settings to factory settings.
Syntax
"∗RST"
Remarks
This command sets the molbox RFM settings to factory settings
which is equivalent to pressing [SPECIAL] on the front panel
and selecting <5Reset>, 1sets. This does not affect the
communications settings.
Example
Command:
Reply:
∗SRE(?)
Purpose
Read or set the service request enable register.
Syntax
"∗SRE n"
"∗SRE?"
Parameters
n:
Remarks
The service request enable register determines which bits of the
status byte can set the MSS bit of the status byte and request
service by asserting the SRQ line of the IEEE-488 interface.
Query
n (0 to 255)
Reply
Example
Command:
Reply:
Command:
Reply:
Page 103
"*OPT?"
"NONE" (no options installed)
"microrange" (microrange option installed)
"*RST"
"OK" (using COM1, no reply if IEEE-488 port)
'0 to 255'. This is the decimal representation of
the bit(s) to enable. To allow the MAV and ESB
bits to assert the SRQ line, the argument would
be 32 + 16 = 48.
Bit 6 (64) is reserved and
cannot be set.
"*SRE 48"
"OK" using COM1, no reply if IEEE-488 port)
"*SRE?"
"48"