UDU(=)
Purpose
Read or set the user definable pressure unit.
Syntax
"UDU"
"UDU=text, coef"
Arguments
text:
coef:
Defaults
"UDU=UNIT, 1"
Example
Command:
Reply:
Errors
ERR# 2:
ERR# 6:
See Also
3.3.2, "PUNIT"
USERCAL(=)
Purpose
Read or set the user flow measurement adjustment.
Syntax
"USERCAL=adder, mult"
"USERCAL"
Defaults
"USERCAL=0, 1"
Arguments
adder:
Mult:
Remarks
The user definable adjustment lets the user modify the measured
flow. The adder and multiplier are separated by a comma. The
adder is in the current flow unit.
Example
Command:
Reply:
Errors
ERR# 6:
See Also
3.3.6, "PCAL(=)
UTEMP
Purpose
Read or set the temperature reference used for the volumetrically
based mass flow unit with user defined reference temperature.
Syntax
"UTEMP=temp"
"UTEMP"
Defaults
"UTEMP=0" (°C)
Arguments
temp:
Remarks
This temperature reference is used by the user defined units.
Example
Command:
Reply:
See Also
3.2.3, 3.2.3.2, and 3.2.3.3, "FUNIT", "TUNIT"
Page 127
The text to use to identify the unit. It can be up to
4 characters wide. This will be the text used to
specify the unit if selecting it remotely, and that
appears on the molbox RFM.
The coefficient to convert Pascal [Pa] to the user
unit. The pressure (in Pa) will be multiplied by this
coefficient before being displayed.
"UDU=Bar2, .00001"
"Bar2, 0.000010"
The text argument is longer than 4 characters.
The coef argument is invalid.
The flow adder in the current flow unit and gas
type (0 to FS).
The flow multiplier (0.1 to 2).
"USERCAL=.58,.995"
"0.58 sccm, 0.995000"
The adder or multiplier argument is invalid.
The
reference
temperature
temperature unit of measure.
"UTEMP=20"
"20.00 C"