Fluke MOLBOX RFM Operation And Maintenance Manual page 131

(ver 1.10 and higher) reference flow monitor
Table of Contents

Advertisement

REMOTE
Purpose
Enable remote local lockout operation.
Syntax
"REMOTE"
Remarks
A REMOTE command deactivates the front panel. All front panel
controls will be disabled. The REMOTE command can only be
canceled by a LOCAL command or by turning OFF the molbox
RFM power then re-applying it.
Example
Command:
Reply:
See Also
"LOCAL"
RES(=)
Purpose
Read or set the local flow display resolution.
Command
"RES=n"
Query
"RES"
Default
"RES=0.01%"
Arguments
n:
Remarks
The molbox's front panel resolution can be specified with this
command. Displayed resolution becomes 10 times smaller than the
current RES setting (ie. default = 0.001% FS) when the microrange
option is active. The resolution of remote flow measurement replies
are always 0.001% FS (0.0001% FS with microrange).
Example
Sent:
Query reply:
Errors
ERR# 6:
See Also
3.4.9
RESET
Purpose
Reset the molbox RFM to the default operating parameters.
Syntax
"RESET"
Remarks
The RESET command can be given to return certain molbox
RFM settings to a default state. This reset corresponds to the
RESET - SETS function from the front panel. The reset process
can take up to 10 seconds to complete, so the user must wait this
time interval before attempting to communicate to the molbox
RFM again.
Example
Command:
Reply:
See Also
3.4.1.7
SN
Purpose
Read the serial number of the molbox RFM.
Syntax
"SN"
Remarks
Each molbox RFM is serialized.
imprinted on the product label on the bottom of the molbox RFM
and is displayed in the power up introductory screen.
Example
Command:
Reply:
See Also
3.4.5.4
Page 121
"REMOTE"
"REMOTE"
Resolution in %FS (1% - 0.01%)
"RES=.01"
"0.010%"
The argument was out of bounds.
"RESET"
"RESET"
This serial number is also
"SN"
"SN"
© 1998-2007 DH Instruments, a Fluke Company
4. REMOTE OPERATION

Advertisement