ERR
Purpose
Read the last error message.
Syntax
"ERR"
Remarks
The "ERR" command provides more details about an error that
has occurred.
"ERR" command returns a brief description about the last error
number that was replied.
Example
Command:
Reply:
See Also
4.3.3
FA=
Purpose
Start a single flow averaging cycle.
Syntax
"FA=period"
Arguments
period:
Remarks
This command is used to start an averaging period that results
in the average molbloc flow and standard deviation of the
molbloc flow over this period. After sending this command you
can monitor the averaging cycle by using the "FR" or "SR"
command to determine when the averaging cycle is complete.
Once the cycle is done, you must use the "FRA" command to
collect the results.
Example
Command:
Reply:
Errors
ERR# 6:
See Also
3.2.6.2, "FR", "SR", "FRA", "ABORT"
FCOEF
Purpose
Read the coefficient that converts kg/s to the current flow units.
Syntax
"FCOEF"
Remarks
The flow coefficient is a value that is used to convert kg/s to the
current flow units. It is dependent on the current flow unit and
the gas selected. To use this coefficient, multiply it by kg/s to
get flow in the current flow units.
Example
Command:
Reply:
See Also
3.2.2. and 7.1.3
Page 109
If the user receives an "ERR# nn" reply, the
"ERR"
"Missing or improper command argument(s)"
The averaging period in seconds (from 4 to 999
seconds).
"FA=20"
"20 s"
The period argument is not within the specified
limits.
"FCOEF"
"4.798073e+004"