Fluke MOLBOX RFM Operation And Maintenance Manual page 124

(ver 1.10 and higher) reference flow monitor
Table of Contents

Advertisement

molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
Errors
See Also
LOCAL
Purpose
Syntax
Remarks
Example
See Also
MEM
Purpose
Syntax
Remarks
Example
MICRORNG(=)
Purpose
Syntax
Defaults
Arguments
Remarks
© 1998-2007 DH Instruments, a Fluke Company
ERR# 6:
One of the arguments is not invalid.
ERR# 15:
A leak check cycle has not been started.
ERR# 23:
The microrange option is not installed in this
RFM.
ERR# 46:
A large leak exists.
ERR# 47:
The temperature changed too much to test for a
ERR# 48:
leak.
ERR# 49:
The absolute RPT disagreement too great to test
ERR# 50:
for a leak.
ERR# 51:
Possible upstream leak.
Possible downstream leak.
Possible bypass leak.
3.2.4.3, "ABORT"
Enable the front panel controls if in remote and go to local mode.
"LOCAL"
In LOCAL mode, all front panel operations are available. The
LOCAL command deactivates REMOTE mode.
Command:
"LOCAL"
Reply:
"LOCAL"
"REMOTE"
Read the memory test status.
"MEM"
On power up, a memory test is run to check the integrity of the
internal data NVRAM.
"FATAL MEMORY FAULT" will be displayed on power up to alert
the user, and the memory test status command will return a '0'.
The command will return a '1' if the memory is OK.
Command:
"MEM"
Reply:
"0"
Read or set the optional microrange operation mode.
"MICRORNG=mode"
"MICRORNG"
"MICRORNG=AUTO"
mode:
"OFF": AUTO and MANUAL MICRORANGE
functions are turned OFF.
"AUTO": AUTO MICRORANGE function is turned
ON.
"ON": MANUAL MICRORANGE function is turned
ON.
The AUTOMATIC and MANUAL MICRORANGE functions control
the way that the microrange option is used in the molbox RFM.
The command in it's query form can also be used to tell if the
microrange option is installed in a molbox RFM.
Page 114
If the memory has been corrupted,