Fluke MOLBOX RFM Operation And Maintenance Manual page 122

(ver 1.10 and higher) reference flow monitor
Table of Contents

Advertisement

molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
ID(=)
Purpose
Syntax
Default
Arguments
Remarks
Example
See Also
KFACT(=)
Purpose
Syntax
Arguments
Defaults
Remarks
Example
See Also
© 1998-2007 DH Instruments, a Fluke Company
Read or set the user defined identification label.
"ID=string"
"ID"
"ID=NONE" (no microrange option)
string:
An alphanumeric string up to 12 characters wide.
The user defined ID label can be used to allow the user to "tag"
the molbox RFM with a unique identifier. This ID is stored in
non-volatile memory and cannot be erased by a power failure,
system fault or reset.
Command:
"ID=RFM-001"
Reply:
"RFM-001"
3.4.5.4
Read or set the molbox RFM K factor.
"KFACT=kfactor"
"KFACT"
kfactor:
The new "K" factor.
"FACT=1.0" (Disabled)
The "K" factor is set to 1 for normal operation.
Command:
"KFACT"
Reply:
"1.000000"
3.2.1
Page 112