8. ELECTRICAL CONNECTIONS
Connect the power cable to the AC inlet socket
(2) provided on the back of the swivel arm as
shown in the fig.-7.
Switch on the power from on/off switch (1).
Note: Power supply is designed with universal
input 100V - 240V AC, 50/60HZ. To plug in follow
Instruction on electrical label provided at bottom
of the arm as shown (3) here in Fig.-7.
8.1 REPLACING THE ILLUMINATION
SOURCE
Open the swivel arm covers. Detach the fibre
optic cable and replace the illumination assembly
A with new assembly. Secure back the arm covers's.

8.2 FUSE REPLACEMENT

The fuse is located with the AC inlet, provided near
the on/off switch. Use a flat head screw driver to
open the fuse compartment. Two fuses are provided
in this, i.e. one is live fuse and second as spare fuse.
Replace the blown fuse with live fuse and secure
back the fuse compartment.
For fuse replacement refer label shown in fig.-9.
8.3 DATA PLATE FOR PRIMA GN MICRO-
SCOPE
Device Name
Model Number
Serial Number
Input Voltage
Frequency
Max. Load
Separate Disposal
waste electric device RX Only Manufacturer
6165000-795
UL Mark
Safety Certification
Accompanying Document
must be consulted
Company
Trademark
Prima GN
2 1
3
Fig. 7
Fig. 8
On/off Switch
AC Inlet
Fig. 9
IEC60601-1:2005
Labeling for USA & Europe
IEC60601-1:2005
Labeling for Canada
Fuse Rating
Label Control
Number
PRIMA GN
A
11
Issue 1.9
Printed on May,2023