GUIDANCE TABLES (continued)
Electromagnetic Emissions
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Prima GN is intended for use in the electromagnetic environment specified below. The customer or
user of the Prima GN should ensure that it is used in such an environment.
IEC 60601 Test
Immunity Test
ESD
± 6kv contact
IEC 61000-4-2
± 8kv Air
EFT
± 2kv Mains
IEC 61000-4-4
± 1kv I/Os
± 1kv Differential
Surge
± 2kv Common
IEC 61000-4-5
Voltage dips,
>95% Dip
dropouts,
for 0.5 Cycle
IEC 61000-4-11
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
Power Frequency
3A/m
50/60Hz Magnetic
Field IEC 61000-4-8
6165000-795
Guidance and Manufacturer's Declaration
All Equipment and Systems
Compliance Level
Level
± 6kv contact
± 8kv Air
± 2kv Mains
± 1kv I/Os
± 1kv Differential
± 2kv Common
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
3A/m
Electromagnetic Environment
Floors should be wood, concrete or ceram-
ic tile. If floors are synthetic, the R/H
should be at least 30%.
Main power quality should be that of a
typical commercial or hospital environ-
ment.
Main power quality should be that of a
typical commercial or hospital environ-
ment.
Main power quality should be that of a
typical commercial or hospital Environ-
ment. If the user of the Prima GN
requires continued operation during
power mains interruption, it is Recom-
mended that the Prima GN be Powered
from an uninterruptable power supply or
battery.
Power frequency magnetic field should be
that of a typical commercial or hospital
Environment.
Prima GN
- Guidance
Issue 1.9
Printed on May,2023
PRIMA GN
27