10. INSTRUCTIONS FOR USING THE COLPOSCOPE
WARNING:
INSTRUMENT IS UNSAFE FOR MRI ENVIRONMENT.
SETTING UP OF COLPOSCOPE:
1. Lock all the brakes on base wheels after setting
up of microscope on the attendance area for
stability.
2. Although tension on microscope is factory
as per the ordered configuration. Still user can
adjust up and down force by tuning the allen
screw clockwise or anticlockwise with the help
of a 8mm allen wrench. Refer Fig.-12 to see the
exact location of tension adjusting screw.
3. Lock the up & Down movement of suspension
arm using locking knob (B) in fig.-12 after
focusing of the area of interest.
4. Adjust the eye distance as per IPD scale
to your convenience.
5. The illumination is controlled through the control
knob (A), fig.-12. Rotate it clockwise
or anticlockwise to achieve desired illumination
level functional setup keep it as minimum.
SETTING UP OF MAGNIFICATION
(Ref. Fig. 13)
1. Adjust to highest magnification with one of the
rotating knobs (11a) provided at magnification
changer.
2. Fine focusing is done by knob provided on CMO
shown as (11d) in fig. 13.
3. Absolute centering of focused area in field of
view can be done by manual handles (11c).
4. Make sure that the magnification changer is
engaged in the index point at the click stop
position.
6165000-795
A
11d
Prima GN
PRIMA GN
B
Tension
adjustment screw
Fig. 12
11a
11c
Fig. 13
14
Issue 1.9
Printed on May,2023