Care And Maintenance - Labomed PRIMA GN User Manual

Colposcope
Hide thumbs Also See for PRIMA GN:
Table of Contents

Advertisement

CARE AND MAINTENANCE

Remove coarse contamination using a damp disposable cloth.
• Use disinfectants based on the following ingredients: aldehyde, alcohols, quaternary ammonium
compounds.
Camera: Clean optical components using a lint-free cloth. Soak the cloth using a little methanol or
glass cleaner. Do not use ethanol and spirit.
• Do not clean products and optical components in a cleaning/disinfecting device or ultra sound bath.
LABOMED MaxiLite coatings are fungal resistant. If you clean as described above, the coatings will
not be damaged.
TROPICAL ENVIRONMENT/FUNGUS:
LABOMED employs certain safety precaution in its manufacturing techniques an and materials. other
preventive measures include:
Keep optical parts clean.
Use and store them in a clean environment only.
Store under UV light when not in use.
Use in continuously climate-controlled rooms only.
Keep moisture away using silica gel and cover with a plastic cover.
OCCUPATIONAL SAFETY AND HEALTH PROTECTION:
Observe work safety and health protection of persons responsible for processing contaminated
products.
Current regulation of hospital hygiene and prevention of infection must be observed in the preparation,
cleaning and disinfection of the products.
INSTRUCTIONS
WORKPLACE:
Remove surface contamination with a paper towel.
REPROCESSING:
Recommended reprocess a product immediately after use or as & when required as per below
cleaning instructions.
CLEANING & SERVICING:
Needed: water, detergent, spirit, microfiber cloth
• Take a Linen or any soft cloth. Moist it Slightly with running tap water (<40°C), using a little
detergent and clean the metallic and plastic parts.
Clean all optical components with spirits or alcohol.
Dry optical components using a microfiber cloth; dry the rest of the product using a paper towel.
For servicing as and when required, inform LABOMED after- sales service department.
AUTOCLAVING:
The rubber caps, sleeves and grips supplied by labomed are recommend for the following program
for autoclaving:
Temperature:
Time:
Instrument:
6165000-795
134° C
10 minute
Standard, Autoclave
Prima GN
PRIMA GN
20
Issue 1.9
Printed on May,2023

Advertisement

Table of Contents
loading

Table of Contents