5.4.35. Identify?
Enables read out the instrument identification.
Example of Use:
IDentify?
enables read out of:
B&K
1050
5.4.36. lnput_Sens
Dummy Message, 1 050 takes no action when read in. Exists so that it can be read back
to the 1 050 without causing an error when accompanying other Messages, e.g. when
reading back the settings, previously read out and stored using the Setup? Message
(section 5.4.61 ).
5.4.37. lnput_Sens?
Enables read out of the input sensitivity, set using the rear panel Input Sensitivity switch
and finely adjusted using the front panel Vibration Meter "Input Sens." function.
Example of Use:
lnput_Sens?
enables read out of, e.g.:
I_S
3 . 96pC/N
5.4.38. Max_Levei_No
- If the rear panel Input Sensitivity switch is set to
"4pCIN", and adjusted to read 3,96pC/N using the
front panel "Input Sens." function.
Selects the number of vibration levels as in the front panel Vibration Meter "Set Levels"
function.
Example of Use:
Max_Levei_No
3
- Sets 3 vibration levels
Notes:
1 to 9 vibration levels can be set if the rear panel Input Sensitivity switch is set to the
"Unit I ms-
2
"
settings, otherwise 1 force level is available if set to "Unit I N".
5.4.39. Max_Levei_No?
Enables read out of the number of vibration levels.
Example of Use:
Max_Levei_No?
enables read out of, e.g.:
M_L_N
7
- For 7 vibration levels