BRUEL & KJAER 1050 Instruction Manual page 16

Vibration exciter control
Table of Contents

Advertisement

LOG DC OUTPUT:
0 to 10V (200mV/dB)
EXTERNAL FILTER
IN/OUT:
8-pin DIN socket for connection to track ing
filter.
Overall Accuracy: 4%
COMPRESSOR:
Dynamic Range: 100 dB
Rectifier: RMS
Regulation Speed:
0,1
to
1000dBs-
1
(in-
creases in proportion to frequency up to the
maximum speed)
Maximum
Speed: 1, 3, 10, 30, 100, 300,
1000dBs-
1
Rate: 0,1 - 0,3
- 1
- 3 dBs-
1
Hz-
1
or OFF
TEST PROGRAMME CONTROL:
Profile: Constant force or constant
acc/vel/disp, programmed in any combina-
tion with up to eight frequency controlled
cross-over
points.
Cross-over Frequencies: Internal calculation
of cross-over frequencies or direct setting.
Cross-over frequency separation must be
>
7% of the lower cross-over frequency.
SAFETY CONTROL:
At the transition to operate, control set-up
compatibility is checked. Error indication is
given for
incompatibility.
Safety Limits: 3,
4, 5,
6dB or OFF. Indepen-
dent setting of upper and lower limit.
System Start-up: Selectable max output lev-
el and attenuated operating level of -20 dB.
System Shut-down: 20 dB per period of fre-
quency
signal.
DIGITAL INTERFACE:
Conforms to IEEE 488 and lEG 625-1 stan-
dards.
Functions Implemented:
Source handshake SH 1
Acceptor handshake AH 1
Talker T 5
Listener L 3
Service request SR 1
Remote local RL 1
Parallel poll PP 0
Device clear DC 1
Device trigger DT 1
Controller C 0
Data Output: Generator
frequency,
generator
level,
meter level and number of elapsed
sweeps.
Remote Control: Input and output of front
panel
set-up (manually or remotely con-
trolled) via the digital interface.
Code: ISO 7-bit code
Status: Besides the service request status
byte,
an error code can be output to give
further information in case of an abnormal
event.
AUXILIARY:
Serial interface used to
control
setting of
auxiliary instruments. When connected these
act as an expansion of the front
panel,
both
manually and remotely
controllable.
RECORDER INTERFACE:
X- Y Recorder: Remote control of pen
up/down and X-deflection (0 to 10V lin or log
de output).
Lin: 0 V at lower sweep limit and 10 V at upper
sweep limit.
·
Log: OV at 1Hz and 10V at 10kHz (2,5V/de-
cade).
Frequency Limitations for Displacement Measurement Range
99,9mm
~
~
' ? _....
IS',>
10mm
'E
Q)
1
mm
E
Q)
(.)
«!
a.
f/)
i:5
~
100
~
m
10~tm
1Hz
10
"
""
~
ooo
100
Frequency
'?
_. . . .
IS',>
""
1kHz
~
10kHz
8518
1711
Frequency Lim itations for Velocity Measurement Range
100mm/sr--------------t--------------~--------------~------------~
Level Recorder: Remote control of start,
stop, automatic stop and pen lift.
POWER REQUIREMENTS:
Power Supply: 100, 115, 127,
200,
220 or
240 V AC
±
20%, 50/60Hz
Consumption: 42 VA approx.
Fuse Requirements: 200 mA (200 to 240
V),
400mA (100 to 127V)
ENVIRONMENTAL:
Safety: Complies with lEG 348 Safety Class 11
Operating Temperature:
+
5 oc to
+
40 oc
( +
41 oF to 1 04 oF)
Storage Temperature: -25 oc
to
+
70 oc
(-13
°
F to 158
°
F)
Humidity Range: 90% RH (non-condensing at
30
°
C)
Electromagnetic Compatibility: Complies
with requirements for Class B computing
de-
vice of American FCC (Federal Communica-
tion Commission) Rules.
GENERAL:
Cabinet: Supplied as model A (lightweight
metal cabinet) or C (as A but with flanges
Frequency
Type KS0023 for 19in racks)
Height: 133 mm (5,2 in)
Width: 430mm (16,9 in)
Depth: 320 mm (12,6 in)
Weight: 10kg (221b)
ACCESSORIES INCLUDED:
8518 16
1 mains cable ......................................... AN 0020
1 UNF to TNC cable .............................. AO
0231
2 spare fuses
.............................. .............
VF 0012
3 spare fuses
...........................................
VF 0039
1
8-
pin DIN plug ...................................... JP0802
1 7 -pin DIN plug
..............................
........ JP 0703
1 6-pin DIN plug
.......................
............... JP0606
3 BNC plugs ............................................. JP 0035
1 IEEE Bus Connection Kit
..................
UA 0814
ACCESSORIES AVAILABLE:
Resonance Dwell Unit... .................... Type 5885
Control Signal Selector .................... Type 5686
Narrow Band Random Noise Unit .. Type 5887
Slave Filter
..........................................
Type 5888
Rack Mounting Flanges
................
......... KS 0023
IEEE 488 interface cable (2m)
....
......... AO 0265
25-pin lEG 625-1 to IEEE 488 in-
terface cable (2m)
.................. ...............
A00264

Advertisement