BRUEL & KJAER 1050 Instruction Manual page 77

Vibration exciter control
Table of Contents

Advertisement

10. Error no. 152.
Error No.
03
11
13
14
15
16
17
18
20
21
22
23
24
25
26
27
28
29
31-38
41-49
50
51
52
53
55
56
57
58
59
60
61
63
This error appears if any of the Resets 1 to 5 is attempted when the
1050 is in "Operate" with the compressor on, or is remotely controlled
with local lock -out.
Description
Source Handshake error (NDAC AND NRFD
=
False, no Listener)
Header code not recognized (not allowed in this instrument)
Text Data received (not allowed in this instrument)
Binary Data received (not allowed in this instrument)
No Output Data exists for this header
This Header is for Output only
Character Data is not allowed for this header
* (asterisk) is not allowed for this Header
Error at argument for: Frequency
Error at argument for: Generator _Levei_Actual or Generator _Levei_Programmed
Error at argument for: Sweep_Rate
Error at argument for: Sweep_Count
Error at argument for: Sweep_Limits
Error at argument for: Compressor _Safety_Limits
Error at argument for: Compressor _Rate
Error at argument for: Compressor _Max_ Speed
Error at argument for: lnput_Sensitivity
Error at argument for: Max_Levei_No.
Error at argument for: Cross_Over_Frequency_N (N
=
1 ... 8)
Error at argument for: Set_Levei_N (N
=
1 ... 9)
Error at argument for: Externai_Filter
Error at argument for: Meter_ Time_ Constant
Error at argument for: Tracking_Filter _Bandwidth
Error at argument for: Narrow_N_Band_Noise
Error at argument for: Recorder _Mode
Error at argument for: Optionai_Parameter
Error at argument for: Meter _Units
Error at argument for: X_Ramp_Sensitivity
Error at argument for: Compressor _Cross_ Over _Safety
Error at argument for: Setup
Error at argument for: Generator _Output
Error at argument for: SWeep
T01097GBO
Table
7.
1. Error Messages
69

Advertisement

Table of Contents