Checking the Developer Temperature
[1]
When the DEVELOPER LAMP begins to flash, remove the TOP COVER.
[2]
Insert a THERMOMETER of known accuracy into the DEVELOPER TANK.
[3]
Install the TOP COVER.
[4]
Operate the processor for 15 minutes to
stabilize
the temperature.
[5]
Check that the temperature on the DEVELOPER TEMPERATURE METER is the same as on the
THERMOMETER (35°C [95°F] set at the factory).
NOTE
See the Service Manual for the adjustment procedure.
[6]
Check that the following are installed:
• RACKS
• CROSSOVERS
• COVERS
[7]
Check that the LIGHT-TIGHT GASKET is tight against the wall.
Checking the Film Transport
[
1
]
Feed three 35 x 43 cm (14 x 17 in.) films into the processor. Check that the RACKS move the film
through the processor correctly.