Kodak X-OMAT M6B Installation Instructions Manual page 33

Hide thumbs Also See for X-OMAT M6B:
Table of Contents

Advertisement

[3]
To obtain the correct static pressure, as shown in Tabie 1, adjust the ciearance at the BUILDING END of
the DUCT as illustrated.
Table 1 Static Pressures
Duct Diameter
Negative Static Pressure, (Water Head)
MIN
MAX
76 mm (3 in.)
0.76 mm (0.03 in.)
1.02 mm (0.04 in.)
102 mm (4 in.)
0.25 mm (0.01 in.)
0.51 mm (0.02 in.)
[4]
Connect the DUCT to the building exhaust.
H108_0008BAA
H108_0008BA
Figure 34 Connecting the Processor Exhaust to the Building Exhaust
[5]
Connect the DUCT to the processor.
246626
33