Charging The Battery - O-Two eAdvantage 01EQ3000 User Manual

Electronic nitrous oxide/oxygen analgesic gas mixing and delivering system
Hide thumbs Also See for eAdvantage 01EQ3000:
Table of Contents

Advertisement

8.2 Charging the Battery

8.2.1 Connect one end of the external power supply/charger to its
supply (100 to 240 Volts or on-board vehicle socket) and the other
end to the DC input socket (C) in Fig. 1 located on the side panel of the
mixer. The indicator lights up as follows:
8.2.2 Turn unit ON and observe battery level (section 1 of the screen).
Refer to 7.3 battery status indicator for exact battery charging status.
The battery shall be fully charged.
Note:
Battery Life expectancy:
Given normal storage and usage, the battery delivers 75% more of its
initial capacity after 300 charge/discharge cycles where the charge
phase is CC/CV 3.3A, 16.80 V and the discharge is 3,3A down to 9000-
mV pack voltage at 25
Battery Shelf life:
The battery provides a minimum of 6 months shelf-life with its initial
charge state of 40%, when stored at 25
The battery pack should be replaced after its life expectancy or if the
battery pack will not fully charge (as indicated on the battery display on the
mixer) or if the mixer doesn't run for more than 5 hours on a single charge.
36
WARNING
The ambient temperature shall be between
0
C and 35
C during battery charge.
o
o
C.
o
WARNING
C.
o
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)