Return to Service
1. Press Reset Timer button on Change Filter screen.
2. Push analyzer into cabinet.
3. Return analyzer to monitor mode in Runtime Options window.
6.3.4 Change Chemcassette™ Tape
Change the Vertex M Chemcassette™ tape for any of the following reasons:
Scheduled end-of-tape service
l
Low Chemcassette™ warning (fault 102)
l
Chemcassette™ has expired (fault 109)
l
End of Chemcassette™ (fault 203)
l
Transport error
l
Preparation
1. Reset End of Chemcassette™ event if present.
2. Touch Main Screen, Menu and then Maintenance.
3. Touch the module number in the Maintenance window.
4. Touch Load CC. Follow the on-screen instructions which will guide you through the
following sequence.
Change Chemcassette™ Tape
1. Pull analyzer out of cabinet.
2. Remove the old Chemcassette™.
3. Install new tape. Verify the RFID tag is aligned with the RF sensor.
4. Touch the NEXT button to read RF sensor.
5. Thread tape leader.
6. Position leading alignment mark on Chemcassette™ tape under front edge of optic head.
7. Touch the NEXT button to verify optics. The tape advances as the Vertex M verifies the
optics.
8. Push the analyzer into the cabinet.